• DocumentCode
    2934342
  • Title

    Calibration of precision measuring amplifiers for strain gauge transducers

  • Author

    Bohacek, J. ; Nakamura, Y. ; Yoshihiro, K. ; Endo, T.

  • Author_Institution
    Fac. of Electr. Eng., Czech Tech. Univ., Prague, Czech Republic
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    120
  • Lastpage
    121
  • Abstract
    A method is described that utilizes a ratio device consisting of multi-decade inductive voltage dividers to calibrate precision measuring amplifiers for strain gauge transducers. Most frequently, this device is used to calibrate 2.5 mV/V ranges of the amplifiers, relative standard uncertainty being 5 ppm for the ratio of 0.0025 and at 225 Hz.
  • Keywords
    bridge circuits; calibration; instrumentation amplifiers; measurement uncertainty; strain gauges; voltage dividers; 225 Hz; calibration; equivalent circuit; multi-decade inductive voltage dividers; precision measuring amplifiers; ratio device; relative standard uncertainty; strain gauge transducers; Admittance; Calibration; Electric variables measurement; Equivalent circuits; Impedance measurement; Laboratories; Measurement standards; Strain measurement; Transducers; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.699813
  • Filename
    699813