DocumentCode
293853
Title
Investigation into the use of position sensitive photomultiplier tubes for beta radiography
Author
Weisenberger, A.G. ; Majewski, S. ; Mehrotra, S. ; Popov, V. ; Saha, M.
Author_Institution
Continuous Electron Beam Accel. Facility, Newport News, VA, USA
Volume
1
fYear
1994
fDate
30 Oct-5 Nov 1994
Firstpage
351
Abstract
Various radioisotopes such as P32 and C14 are used as radiolabels for autoradiography in biological and gene research. We have investigated the possibility of using two types of photomultiplier tubes coupled to plastic and crystal scintillators to detect and quantify the beta emissions of P32 and C14 in amounts typical for autoradiography of dot-blot bioassays in the microtiter and other formats. The Hamamatsu R4760 (with 16 individual channels) and R2486 (a position sensitive PMT with 16×16 cross-wire anodes) were the two multichannel photomultiplier tubes tested. The levels of beta radiation typically found in autoradiographic experiments are shown to be detectable using plastic and crystal scintillators coupled to these photomultiplier tubes
Keywords
beta-ray detection; detector circuits; diagnostic radiography; nuclear electronics; photomultipliers; position sensitive particle detectors; solid scintillation detectors; 16×16 cross-wire anodes; C14; Hamamatsu R4760; P32; R2486; autoradiographic experiments; autoradiography; beta emissions; beta radiography; crystal scintillators; microtiter; multichannel photomultiplier tubes; photomultiplier tubes; plastic scintillators; position sensitive PMT; position sensitive photomultiplier tubes; radioisotopes; radiolabels; Anodes; DNA; Phosphors; Photomultipliers; Plastics; Radiation detectors; Radioactive materials; Radiography; Testing; Wire;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record
Conference_Location
Norfolk, VA
Print_ISBN
0-7803-2544-3
Type
conf
DOI
10.1109/NSSMIC.1994.474363
Filename
474363