• DocumentCode
    2939277
  • Title

    Preliminary Test of Steam Humidity Measurement with Capacitance Sensor

  • Author

    Tian Rui-feng ; Zhang Peng-fei ; Du Li-peng ; Liu Xiao-yi

  • Author_Institution
    Nat. Defense Key Discipline Lab. of Nucl. Safety & Simulation Technol., Harbin Eng. Univ., Harbin, China
  • fYear
    2011
  • fDate
    25-28 March 2011
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    Applying the theory of capacitance measurement and capacitor working principle, a kind of capacitive steam humidity measurement system has been developed, which includes steam humidity adjusting system, external measuring circuit and steam humidity calibration system. Use this steam humidity measurement system; a kind of coaxial cylindrical capacitance sensor is studied. The results shows that, the coaxial cylindrical capacitance sensor has stable static characteristics; capacitance sensor output characteristic does not change with the steam flow; the output capacitance of the sensor increases with the steam humidity, and they has a good linear relationship.
  • Keywords
    capacitance measurement; capacitive sensors; humidity measurement; steam turbines; capacitance measurement; capacitor working principle; coaxial cylindrical capacitance sensor; external measuring circuit; steam flow; steam humidity calibration; steam humidity measurement; Capacitance; Capacitance measurement; Dielectric constant; Humidity; Humidity measurement; Turbines; Water heating;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Power and Energy Engineering Conference (APPEEC), 2011 Asia-Pacific
  • Conference_Location
    Wuhan
  • ISSN
    2157-4839
  • Print_ISBN
    978-1-4244-6253-7
  • Type

    conf

  • DOI
    10.1109/APPEEC.2011.5749007
  • Filename
    5749007