Title :
Radionuclide mapping using a Position Sensitive Photomultiplier Tube and a germanium detector
Author :
Dewaraja, Yuni K. ; He, Zhong ; Fleming, Ronald F. ; Wehe, David K. ; Fleming, Ronald H.
Author_Institution :
Dept. of Nucl. Eng., Michigan Univ., Ann Arbor, MI, USA
fDate :
30 Oct-5 Nov 1994
Abstract :
This paper demonstrates simultaneous two-dimensional mapping of multiple radionuclides which are present in a sample. The technique, which is based on position sensitive beta-gamma coincidence measurement, is applicable to most radionuclides that decay by beta emission followed by the emission of a coincident gamma-ray. In the present work, a Position Sensitive Photomultiplier Tube is used for imaging the beta particles while a Ge detector is used to identify the radionuclides by detecting the coincident gamma-rays
Keywords :
beta-ray detection; coincidence techniques; detector circuits; gamma-ray detection; germanium radiation detectors; nuclear electronics; photomultipliers; position sensitive particle detectors; Ge detector; Position Sensitive Photomultiplier Tube; beta emission decay; coincident gamma-ray emission; multiple radionuclides; position sensitive beta-gamma coincidence measurement; radionuclide decay; radionuclide mapping; two-dimensional mapping; Beta rays; Circuits; Gamma ray detection; Gamma ray detectors; Germanium; Optical imaging; Photomultipliers; Position measurement; Position sensitive particle detectors; Simultaneous localization and mapping;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record
Conference_Location :
Norfolk, VA
Print_ISBN :
0-7803-2544-3
DOI :
10.1109/NSSMIC.1994.474480