Performance of the BGO endcap calorimeter with phototriode readout for the CMD-2 detector
Author
Grigoriev, D.N. ; Akhmetshin, R.R. ; Beschastnov, P.M. ; Fedorenko, V.E. ; Pyata, E.E. ; Smakhtin, V.P. ; Yudin, Yu V. ; Shlegel, V.N. ; Vasiliev, Ya V.
Author_Institution
Budker Inst. of Nucl. Phys., Novosibirsk, Russia
Volume
2
fYear
1994
fDate
30 Oct-5 Nov 1994
Firstpage
533
Abstract
The endcap calorimeter of the CMD-2 detector for the VEPP-2M collider is described. The calorimeter contains 680 BGO crystals read out by the vacuum phototriodes. Charge sensitive preamplifiers are placed directly on the phototriodes for the best noise performance. The calorimeter was installed during the last year. Calibration by cosmic rays showed the energy equivalent of the electronic noise to be about 0.9 MeV during operation in a 1 T magnetic field which corresponds to design value
Keywords
calibration; calorimeters; detector circuits; electron device noise; nuclear electronics; phototubes; solid scintillation detectors; triodes; 1 T; BGO; Bi4Ge312; CMD-2 detector; calibration; charge sensitive preamplifiers; cosmic rays; endcap calorimeter; noise; phototriode readout; Energy resolution; Magnetic fields; Mesons; Notice of Violation; Nuclear physics; Photonic crystals; Region 8; Solid scintillation detectors; Stability; Tiles;