Title :
Cardiac output and stroke volume estimation using a hybrid of three Windkessel models
Author :
Arai, Tatsuya ; Lee, Kichang ; Cohen, Richard J.
Author_Institution :
Dept. of Aeronaut. & Astronaut., Massachusetts Inst. of Technol. (MIT), Cambridge, MA, USA
fDate :
Aug. 31 2010-Sept. 4 2010
Abstract :
Cardiac output (CO) and stroke volume (SV) are the key hemodynamic parameters to be monitored and assessed in ambulatory and critically ill patients. The purpose of this study was to introduce and validate a new algorithm to continuously estimate, within a proportionality constant, CO and SV by means of mathematical analysis of peripheral arterial blood pressure (ABP) waveforms. The algorithm combines three variants of the Windkessel model. Input parameters to the algorithm are the end-diastolic pressure, mean arterial pressures, inter-beat interval, and the time interval from end-diastolic to peak systolic pressure. The SV estimates from the three variants of the Windkessel model were weighted and integrated to provide beat-to-beat SV estimation. In order to validate the new algorithm, the estimated CO and SV were compared to those obtained through surgically implanted Transonic™ aortic flow probes placed around the aortic roots of six Yorkshire swine. Overall, estimation errors in CO and SV derived from radial ABP were 10.1% and 14.5% respectively, and 12.7% and 16.5% from femoral ABP. The new algorithm demonstrated statistically significant improvement in SV estimation compared with previous methods.
Keywords :
blood pressure measurement; blood vessels; cardiovascular system; haemodynamics; medical diagnostic computing; physiological models; Transonic™ aortic flow probes; Windkessel model; Yorkshire swine; cardiac output; end-diastolic pressure; estimation errors; hemodynamic parameters; inter-beat interval; mean arterial pressure; peripheral arterial blood pressure; stroke volume; Algorithm design and analysis; Biomedical monitoring; Blood pressure; Estimation; Extraterrestrial measurements; Monitoring; Time measurement; Blood Pressure; Cardiac Output; Computer Simulation; Diagnosis, Computer-Assisted; Humans; Models, Cardiovascular; Reproducibility of Results; Sensitivity and Specificity; Stroke Volume;
Conference_Titel :
Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
Conference_Location :
Buenos Aires
Print_ISBN :
978-1-4244-4123-5
DOI :
10.1109/IEMBS.2010.5627225