Investigation of the count rate performance of the General Electric Advance positron emission tomograph
Author :
Lewellen, T.K. ; Kohlmyer, S.G. ; Miyaoka, R.S. ; Schubert, S. ; Stearns, C.W.
Author_Institution :
Washington Univ., WA, USA
fDate :
30 Oct-5 Nov 1994
Abstract :
A series of 2D and 3D count rate measurements have been carried out on the Advance PET system. The measurements include: NEMA standard protocols; simultaneous imaging of two small vials containing [O-15] water and [F-18]; a cardiac insert placed in a water filled elliptical tank; two different head sized cylindrical tanks; and a Hoffman 3D phantom. The data were processed with and without dead time and scatter corrections. Plots were made of the total count rate, randoms, trues+scatter, scatter corrected trues, and noise equivalent count rate as a function of total activity in the machine and as the μCi/cc in the phantoms. The accuracy of deadtime corrections was also measured. In 2D mode, the NEMA standard protocol resulted in maximum `trues´ count rates of 540,000 cps in high sensitivity mode. In 3D mode the head sized cylindrical phantoms yielded a maximum trues rate of approximately 950,000 cps. Recovery of the water half-life was accurate with up to 50 mCi for the cardiac insert in 2D mode
Keywords :
biomedical equipment; positron emission tomography; 2D count rate measurements; 3D count rate measurements; General Electric Advance positron emission tomograph; count rate performance; ~O-1; Biomedical imaging; Collimators; Image reconstruction; Imaging phantoms; Magnetic heads; Positron emission tomography; Protocols; Radioactive decay; Scattering; Workstations;
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference, 1994., 1994 IEEE Conference Record
Conference_Location :
Norfolk, VA
Print_ISBN :
0-7803-2544-3
DOI :
10.1109/NSSMIC.1994.474637