Abstract
A feasibility study for a high resolution positron emission tomograph, based on 9.5×4×4 mm3 LSO crystals viewed by 3 mm diameter avalanche photodiodes, has been carried out. Using a Na22 source the authors determined a spatial resolution of 2.3±0.1 mm, an energy resolution around 15% and a time resolution of 2.6 nsec. Possible configurations for larger scale tests and a tomograph are given
Keywords
avalanche photodiodes; biomedical electronics; positron emission tomography; 2.6 ns; 3 mm; LSO crystals; Na; Na22 source; avalanche photodiode readout; diagnostic nuclear medicine; energy resolution; high resolution PET; medical instrumentation; nsec time resolution; spatial resolution; time resolution; Animals; Avalanche photodiodes; Crystalline materials; Crystals; Energy resolution; High-resolution imaging; In vivo; Positron emission tomography; Spatial resolution; Testing;