DocumentCode
2944518
Title
Full coverage millimetre wave primary noise standards for 18-170 GHz
Author
Vizard, D.R. ; Foster, P.R. ; Lunn, B. ; Cherry, S.M.
Author_Institution
VivaTech LLC, USA
fYear
2008
fDate
15-20 June 2008
Firstpage
1537
Lastpage
1540
Abstract
Conventional noise diode transfer standards used in the calibration of sensitive radiometers and other low noise components or instruments can produce large errors due to limitations of frequency coverage and absolute accuracy. The Cryogenic Noise Calibration Standard described features interchangeable waveguide based horn antennas allowing full frequency coverage from 18-170 GHz and beyond. The state-of-art design of the vessel and its temperature profile allows the effects of stray radiation to be controlled within the required accuracy specifications.
Keywords
Calibration; Cryogenics; Diodes; Frequency; Instruments; Nitrogen; Radiometers; Standards development; Temperature control; Temperature sensors;
fLanguage
English
Publisher
ieee
Conference_Titel
Microwave Symposium Digest, 2008 IEEE MTT-S International
Conference_Location
Atlanta, GA, USA
ISSN
0149-645X
Print_ISBN
978-1-4244-1780-3
Electronic_ISBN
0149-645X
Type
conf
DOI
10.1109/MWSYM.2008.4633074
Filename
4633074
Link To Document