Abstract
This paper presents the prototype of a new remote guiding device to find the accurate screw hole position for drilling and direction for locking in intramedullary nailing surgery. The device consists of three magnetic pins and an electrical conductive board. The three symmetrically placed magnetic pins can rotate freely and point to the permanent magnet inside the nail. The drilling position and screwing direction can be obtained by moving the device with those magnetic pins to align with the magnet in the nail. The alignment condition is indicated by a LED. When the device is not aligned with the magnet in the nail properly, either in position or direction, the pins then will contact with the board to trigger a light-emitting diode for alarming.
Keywords
biomedical equipment; electrical conductivity; fasteners; light emitting diodes; permanent magnets; surgery; LED; alarm; distal locking; drilling; electrical conductive board; intramedullary nailing surgery; light emitting diode; magnetic pins; permanent magnet; remote guiding device; screw hole position prediction; Bones; Light emitting diodes; Magnetic devices; Magnetic flux; Nails; Permanent magnets; Pins;
fLanguage
English
Publisher
ieee
Conference_Titel
Sensors, 2012 IEEE
Conference_Location
Taipei
ISSN
1930-0395
Print_ISBN
978-1-4577-1766-6
Electronic_ISBN
1930-0395
Type
conf
DOI
10.1109/ICSENS.2012.6411140
Filename
6411140
Link To Document