• DocumentCode
    2951223
  • Title

    High-performance trapped Hg/sup +/ frequency standards

  • Author

    Berkeland, D.J. ; Miller, J.D. ; Young, B.C. ; Bergquist, J.C. ; Itano, W.M. ; Wineland, D.J.

  • Author_Institution
    Time & Frequency Div., Nat. Inst. of Stand. & Technol., Boulder, CO, USA
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    321
  • Lastpage
    322
  • Abstract
    We have demonstrated a new frequency standard, accurate to 3.4/spl times/10/sup -15/, based on the 40.5 GHz ground state hyperfine transition in cooled and trapped /sup 199/Hg/sup +/. We are also developing an optical frequency standard based on /sup 199/Hg/sup +/. The spectral width of the 563 nm laser used in this standard is less than 6 Hz.
  • Keywords
    Zeeman effect; atomic clocks; frequency stability; frequency standards; ground states; hyperfine structure; laser cooling; measurement uncertainty; mercury (metal); microwave measurement; particle traps; positive ions; radiation pressure; trapped ions; 282 nm; 40.5 GHz; Hg; Paul-Straubel trap; Ramsey spectrum; Zeeman shift; clock transition; cooled trapped /sup 199/Hg/sup +/; electric dipole transitions; frequency stability; ground state hyperfine transition; high-performance frequency standard; ion crystals; laser cooling; linear Paul trap; microwave frequency standard; optical frequency standard; spectral width; trapped ions; uncertainty; Acoustic beams; Frequency synthesizers; Laser beams; Laser stability; Laser transitions; Mercury (metals); Optical scattering; Particle beam optics; Resonance; Resonant frequency;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.699930
  • Filename
    699930