• DocumentCode
    2953014
  • Title

    Towards a cheaper, simpler quantized Hall resistance standard

  • Author

    Inglis, A.D.

  • Author_Institution
    Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, Ont., Canada
  • fYear
    1998
  • fDate
    6-10 July 1998
  • Firstpage
    337
  • Lastpage
    338
  • Abstract
    We have produced and characterised devices suitable for accurate quantum Hall resistor (QHR) measurements at fields below 8 T, and temperatures of 1.2-1.5 K. We have made accurate comparisons of QHR devices with wirewound resistors using commercially available room-temperature dc current-comparator bridges. These two results imply that it is possible to assemble an accurate, but relatively cheap and simple to operate, QHR standard.
  • Keywords
    bridge circuits; current comparators; electric resistance measurement; measurement standards; quantum Hall effect; resistors; 1.2 to 1.5 K; 7.5 T; DC current-comparator bridges; accurate comparisons; accurate quantum Hall resistor measurements; commercial bridges; simpler quantized Hall resistance standard; uncertainty; wirewound resistors; Cryogenics; Current measurement; Electrical resistance measurement; Laboratories; Measurement standards; Resistors; Standards development; Superconducting magnets; Temperature; Tin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements Digest, 1998 Conference on
  • Conference_Location
    Washington, DC, USA
  • Print_ISBN
    0-7803-5018-9
  • Type

    conf

  • DOI
    10.1109/CPEM.1998.699939
  • Filename
    699939