DocumentCode :
2954743
Title :
Two international comparisons of SMU 10 V Josephson array voltage standard
Author :
Vrabcek, P.
Author_Institution :
Inst. of Metrol., Bratislava, Slovakia
fYear :
1998
fDate :
6-10 July 1998
Firstpage :
355
Lastpage :
356
Abstract :
Interlaboratory comparisons of the SMU 10 V voltage standard based on Josephson effect have been carried out with the BNM/LCIE and the NPL 10 V Josephson standards. The comparison method was indirect using Zener diode references as the travelling standards. The results show agreement between the SMU and the BNM/LCIE calibrations of the Fluke 732 A reference standard 23.5 nV and between the SMU and the NPL calibrations of the Fluke 732 B 55 nV.
Keywords :
calibration; measurement standards; superconducting junction devices; voltage measurement; 10 V; BNM/LCIE; Fluke 732 A; Fluke 732 B; Josephson array voltage standard; NPL; SMU; Zener diode reference; calibration; international comparison; travelling standard; Calibration; Diodes; Frequency; Josephson effect; Laboratories; Measurement standards; Semiconductor device measurement; Switches; Voltage; Voltmeters;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements Digest, 1998 Conference on
Conference_Location :
Washington, DC, USA
Print_ISBN :
0-7803-5018-9
Type :
conf
DOI :
10.1109/CPEM.1998.699948
Filename :
699948
Link To Document :
بازگشت