Title :
Two international comparisons of SMU 10 V Josephson array voltage standard
Author_Institution :
Inst. of Metrol., Bratislava, Slovakia
Abstract :
Interlaboratory comparisons of the SMU 10 V voltage standard based on Josephson effect have been carried out with the BNM/LCIE and the NPL 10 V Josephson standards. The comparison method was indirect using Zener diode references as the travelling standards. The results show agreement between the SMU and the BNM/LCIE calibrations of the Fluke 732 A reference standard 23.5 nV and between the SMU and the NPL calibrations of the Fluke 732 B 55 nV.
Keywords :
calibration; measurement standards; superconducting junction devices; voltage measurement; 10 V; BNM/LCIE; Fluke 732 A; Fluke 732 B; Josephson array voltage standard; NPL; SMU; Zener diode reference; calibration; international comparison; travelling standard; Calibration; Diodes; Frequency; Josephson effect; Laboratories; Measurement standards; Semiconductor device measurement; Switches; Voltage; Voltmeters;