DocumentCode
2966354
Title
Aortic pressure estimation for the Penn State electric ventricular assist device
Author
Koontz, J.W. ; Tsach, U. ; Geselowitz, D.B.
Author_Institution
Pennsylvania State Univ., University Park, PA, USA
fYear
1988
fDate
10-11 Mar 1988
Firstpage
166
Lastpage
170
Abstract
A computer simulation of the Penn State rollerscrew electric ventricular assist device (EVAD) and the mock circulatory system has been developed for the purpose of control system design. The system model and simulation were verified for accuracy with a mock circulatory system using an EVAD. An observer based on a Luenberger observer was used to predict unmeasurable system pressures necessary for pump output control. From the simulation, accurate aortic pressure values were obtained using the observer during systole at a beat rate of 60 beats/min. The gains used in the observer were judiciously chosen, but a method for optimizing the feedback gains was developed. This method of selecting the observer gains which minimizes the error between the observer and the system is by pole placement rather than by statistical techniques
Keywords
artificial organs; cardiology; haemodynamics; Luenberger observer; Penn State electric ventricular assist device; computer simulation; feedback gain optimisation; mock circulatory system; pole placement; pump output control; systole; Brushless DC motors; Control systems; DC motors; Fasteners; Hall effect devices; Micromotors; Observers; Pressure measurement; State estimation; Torque;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 1988., Proceedings of the 1988 Fourteenth Annual Northeast
Conference_Location
Durham, NH
Type
conf
DOI
10.1109/NEBC.1988.19375
Filename
19375
Link To Document