    Enhancement of pulse contour analysis in the pulmonary artery by use of heart sounds

  • Author

    Kirchner, Jens ; Van Ooyen, André ; Ershov, Sergey ; Skerl, Olaf

  • Author_Institution
    Biotronik SE & Co. KG, Erlangen, Germany
  • fYear
    2011
  • fDate
    28-31 Oct. 2011
  • Firstpage
    1792
  • Lastpage
    1795
  • Abstract
    Pulse contour analysis (PCA) estimates cardiac stroke volume on a beat-to-beat basis from the systolic time interval of arterial blood pressure. Hence, identification of the ejection time intervals plays a crucial role. However, this is an intricate task when pulmonary artery pressure (PAP) measured with an implantable pressure sensing device is used. The feature marking the end of systole is rather indistinct and thus often not detectable in low-resolution signals. We therefore suggest using additional information from heart sounds, which are indicative of beginning and end of systole, respectively. The potential benefit of this approach was evaluated with data from an animal experiment. Blood pressure was recorded with a pressure transducer catheter and heart sounds with a prototype accelerometer. Using PAP only, 6.1% of all cycles were unusable for PCA. With additional use of heart sounds, however, a detection rate of 100.0% was achieved.
  • Keywords
    blood pressure measurement; blood vessels; catheters; pressure sensors; pressure transducers; prosthetics; PAP measurement; PCA enhancement; arterial blood pressure systolic time interval; cardiac stroke volume estimation; ejection time intervals identification; heart sound; implantable pressure sensing device; pressure transducer catheter; prototype accelerometer; pulmonary artery pressure measurement; pulse contour analysis enhancement; Accelerometers; Arteries; Biomedical monitoring; Blood pressure; Heart; Principal component analysis; Sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2011 IEEE
  • Conference_Location
    Limerick
  • ISSN
    1930-0395
  • Print_ISBN
    978-1-4244-9290-9
  • Type

    conf

  • DOI
    10.1109/ICSENS.2011.6127104
  • Filename
    6127104