DocumentCode
2968727
Title
High frequency impedance measurements from implanted unipolar pacing leads
Author
Steinhaus, Bruce M. ; Molan, J.A. ; Samuelson, Kent E. ; Morris, Robert A. ; Dawson, Albert K.
Author_Institution
Telec, Englewood, CO, USA
Volume
1
fYear
1995
fDate
20-25 Sep 1995
Firstpage
333
Abstract
Impedance measurements using frequencies up to 500 MHz were used along with time domain reflectometry techniques to investigate respiration sensing using standard unipolar pacemaker leads. The results have demonstrated the unique nature of these impedance sensing methods versus traditional lower frequency impedance methods. Experimental evidence includes: separation in frequency (or time and thus distance along the lead) of the cardiac stroke volume signal from the respiration signal and respiration sensing using a fully insulated tip unipolar pacing lead
Keywords
bioelectric phenomena; biomedical measurement; electric impedance measurement; pacemakers; patient monitoring; pneumodynamics; time-domain reflectometry; 500 MHz; cardiac stroke volume signal; fully insulated tip unipolar pacing lead; high frequency impedance measurements; impedance sensing methods; implantable rate responsive pacemakers; implanted unipolar pacing leads; respiration sensing; respiration signal; standard unipolar pacemaker leads; time domain reflectometry; Coaxial cables; Fixtures; Frequency measurement; Impedance measurement; Instruments; Measurement standards; Pacemakers; Pulse measurements; Reflectometry; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location
Montreal, Que.
Print_ISBN
0-7803-2475-7
Type
conf
DOI
10.1109/IEMBS.1995.575136
Filename
575136