• DocumentCode
    2973587
  • Title

    Spectral measurement with a UV linear-variable optical filter microspectrometer

  • Author

    Emadi, Arvin

  • Author_Institution
    Dept. ME/EI, Delft Univ. of Technol., Delft, Netherlands
  • fYear
    2011
  • fDate
    28-31 Oct. 2011
  • Firstpage
    420
  • Lastpage
    423
  • Abstract
    This paper presents the design, fabrication and spectral measurements of an Ultra-Violet (UV) Linear Variable Optical Filter (LVOF)-based micro-spectrometer operating in the 300 nm - 400 nm wavelength range. The UV LVOF has been fabricated in an IC-Compatible process using resist reflow. Characterization of the LVOF, by passing monochromatic light through the LVOF, shows high linearity of the profile. The filter provides the possibility to have a robust high-resolution microspectrometer in the UV on a CMOS chip. Spectrum of a Mercury lamp has been measured using the UV LVOF-microspectrometer with 0.5 nm spectral resolution.
  • Keywords
    optical design techniques; optical fabrication; optical filters; ultraviolet spectrometers; UV linear-variable optical filter microspectrometer; resist reflow; spectral measurement; CMOS integrated circuits; Cameras; Cavity resonators; Fabrication; Hafnium compounds; Optical filters; Wavelength measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2011 IEEE
  • Conference_Location
    Limerick
  • ISSN
    1930-0395
  • Print_ISBN
    978-1-4244-9290-9
  • Type

    conf

  • DOI
    10.1109/ICSENS.2011.6127349
  • Filename
    6127349