• DocumentCode
    2974654
  • Title

    High speed cell stiffness evaluation toward 100% reliability

  • Author

    Hirose, Yuki ; Kaneko, Makoto ; Kawahara, Tomohiro ; Yamanishi, Yoko ; Arai, Fumihito

  • Author_Institution
    Sch. of Mech. Eng., Osaka Univ., Suita, Japan
  • fYear
    2011
  • fDate
    28-31 Oct. 2011
  • Firstpage
    962
  • Lastpage
    965
  • Abstract
    This paper proposes a high speed cell stiffness evaluation with high reliability. The procedure includes both online measurement and offline inspection. By using the property where the passing time of the cell in the micro channel is a function of the cell stiffness and its size, a real time vision system whose handling frequency is 6 kHz can capture the size as well as the passing time of each cell. During the online measurement, we save two image data at both the entrance and the exit for each cell. In offline manual inspection, all doubtful data are killed so that we can make sure the reliability. Experimental results ensure that the proposed approach can handle 500 cells within 10min in including both online and offline procedures.
  • Keywords
    microchannel flow; microsensors; reliability; doubtful data; high reliability; high speed cell stiffness evaluation; image data; microchannel; offline inspection; online measurement; real time vision system; Brightness; Cameras; Force; Red blood cells; Reliability; Size measurement; Time measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2011 IEEE
  • Conference_Location
    Limerick
  • ISSN
    1930-0395
  • Print_ISBN
    978-1-4244-9290-9
  • Type

    conf

  • DOI
    10.1109/ICSENS.2011.6127399
  • Filename
    6127399