DocumentCode
2976669
Title
Medical transducer arrays using composite materials for acoustic matching layers
Author
Kondo, Toshiaki
Author_Institution
Fac. of Eng., Tokushima Bunri Univ., Kagawa, Japan
Volume
3
fYear
2004
fDate
23-27 Aug. 2004
Firstpage
1659
Abstract
It is proposed that unidirectional carbon composite materials allow the realization of the materials under 0.1 Poisson ratio. This study is concerned with the manufacturing process of the composite materials adopted for the matching layers of medical transducer arrays and the experimental results on their physical properties. And it is explained by computer simulation and experiment that the use of the composite materials for the matching layers decreases sharply the mechanical coupling between the adjacent elements of the array.
Keywords
Poisson ratio; biomedical transducers; biomedical ultrasonics; composite materials; ultrasonic transducer arrays; Poisson ratio; acoustic matching layers; adjacent elements; composite materials; manufacturing process; mechanical coupling; medical transducer arrays; unidirectional carbon composite materials; Acoustic arrays; Acoustic transducers; Biomedical acoustics; Biomedical transducers; Composite materials; Computer simulation; Impedance; Manufacturing processes; Medical diagnostic imaging; Organic materials;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2004 IEEE
ISSN
1051-0117
Print_ISBN
0-7803-8412-1
Type
conf
DOI
10.1109/ULTSYM.2004.1418141
Filename
1418141
Link To Document