• DocumentCode
    2976669
  • Title

    Medical transducer arrays using composite materials for acoustic matching layers

  • Author

    Kondo, Toshiaki

  • Author_Institution
    Fac. of Eng., Tokushima Bunri Univ., Kagawa, Japan
  • Volume
    3
  • fYear
    2004
  • fDate
    23-27 Aug. 2004
  • Firstpage
    1659
  • Abstract
    It is proposed that unidirectional carbon composite materials allow the realization of the materials under 0.1 Poisson ratio. This study is concerned with the manufacturing process of the composite materials adopted for the matching layers of medical transducer arrays and the experimental results on their physical properties. And it is explained by computer simulation and experiment that the use of the composite materials for the matching layers decreases sharply the mechanical coupling between the adjacent elements of the array.
  • Keywords
    Poisson ratio; biomedical transducers; biomedical ultrasonics; composite materials; ultrasonic transducer arrays; Poisson ratio; acoustic matching layers; adjacent elements; composite materials; manufacturing process; mechanical coupling; medical transducer arrays; unidirectional carbon composite materials; Acoustic arrays; Acoustic transducers; Biomedical acoustics; Biomedical transducers; Composite materials; Computer simulation; Impedance; Manufacturing processes; Medical diagnostic imaging; Organic materials;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2004 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-8412-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2004.1418141
  • Filename
    1418141