Title :
A new performance analysis metric for medical ultrasound
Author :
Ranganathan, Karthik ; Walker, William F.
Author_Institution :
Dept. of Biomed. Eng., Virginia Univ., Charlottesville, VA, USA
Volume :
3
fYear :
2004
fDate :
23-27 Aug. 2004
Firstpage :
1737
Abstract :
The resolution of medical ultrasound systems is usually characterized by the width of the mainlobe, typically the full width at half maximum (FWHM), in beamplots. The FWHM has, however, been shown to sometimes provide misleading information about systems. The concept of "cystic resolution" in which performance is quantified as the size of the void that produces a specified contrast, has been previously introduced to medical ultrasound. We extend the concept and develop a general metric to analyze the 3D broadband performance of arbitrary ultrasound systems. We provide an example of application in which we compare the performance of a conventional system using a 1D array and time delay beamforming of radiofrequency (RF) data to that of a system based on a 2D array that focuses solely via phase rotation of complex I/Q data formed by directly sampling the received RF signal.
Keywords :
array signal processing; biomedical transducers; biomedical ultrasonics; image resolution; signal sampling; ultrasonic imaging; ultrasonic transducer arrays; 1D array; 2D array; 3D broadband performance; RF signal sampling; complex I/Q data; cystic resolution; medical ultrasound; performance analysis metric; phase rotation; time delay beamforming; Biomedical engineering; Biomedical imaging; Performance analysis; Phased arrays; Radar imaging; Radar scattering; Radio frequency; Signal resolution; Ultrasonic imaging; White noise;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2004 IEEE
ISSN :
1051-0117
Print_ISBN :
0-7803-8412-1
Type :
conf
DOI :
10.1109/ULTSYM.2004.1418161
Filename :
1418161
Link To Document :
بازگشت