Title
Volume flow measurements with a novel semi-automated 4D Doppler ultrasound scanner [blood flow measurement]
Author
Forsberg, F. ; Liu, J.B. ; Stein, Aaron ; Deng, Xinhuan ; Abend, Kenneth
Author_Institution
Dept. of Radiol., Thomas Jefferson Univ. Hosp., Philadelphia, PA, USA
Volume
3
fYear
2004
fDate
23-27 Aug. 2004
Firstpage
2101
Abstract
The Encore PV is a novel, real time 3D (i.e., 4D) ultrasound scanner capable of measuring blood flow. In vitro and in vivo volume flow measurements obtained from the Encore were compared to those from an invasive transit time flowmeter. A flow pump was used to simulate carotid type waveforms with pulsatile flow rates from 60 to 600 ml/min. Encore volume flows were compared to an invasive technique for volumetric blood flow measurement utilizing a transit time flowmeter. In vivo 10 s data sets of the volume flow in the distal aorta of 6 rabbits were obtained simultaneously with the Encore PV and the flowmeter. Mean and maximum volume flows were recorded with both units. Data was compared using linear regression and Bland-Altman analysis. It was found that volume flow measurements with this unit are comparable to those of an invasive flowmeter.
Keywords
Doppler measurement; biomedical ultrasonics; blood flow measurement; flowmeters; pulsatile flow; regression analysis; 10 s; Bland-Altman analysis; Encore PV; carotid type waveforms; flow pump; invasive transit time flowmeter; linear regression; pulsatile flow; rabbit distal aorta; real time 3D scanner; semi-automated 4D Doppler ultrasound scanner; volume blood flow measurements; Blood flow; Fluid flow measurement; In vitro; In vivo; Linear regression; Rabbits; Time measurement; Ultrasonic imaging; Ultrasonic variables measurement; Volume measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2004 IEEE
ISSN
1051-0117
Print_ISBN
0-7803-8412-1
Type
conf
DOI
10.1109/ULTSYM.2004.1418251
Filename
1418251
Link To Document