Title :
The electromagnetic environment due to portable sources in a typical hospital room
Author :
Vlach, P. ; Liu-Hinz, C. ; Segal, B. ; Skulic, B. ; Pavlasek, T.
Author_Institution :
Dept. of Electr. Eng., McGill Univ., Montreal, Que., Canada
Abstract :
In order to characterize the electromagnetic environment, due to portable sources, inside a typical Montreal area hospital room, fields were measured in the 148-174 MHz, 425-480 MHz and 825-850 MHz ranges over a period of 24 hours. Call durations in each frequency range remained below 8 seconds 50% of the time, while field levels were 63 dBμV/m (1.4 V/m) or lower, 50% of the time. Similar field monitoring in high traffic areas, such as emergency, for longer periods of time, should enable the characterization of call profiles, thereby leading to improved susceptibility testing of medical equipment
Keywords :
electromagnetic compatibility; electromagnetic interference; microwave measurement; 148 to 174 MHz; 24 hr; 425 to 480 MHz; 8 s; 825 to 850 MHz; call durations; call profiles characterization; electromagnetic environment; emergency rooms; high traffic areas; medical equipment susceptibility testing; portable sources; typical Montreal area hospital room; Biomedical engineering; Biomedical equipment; Biomedical monitoring; Computerized monitoring; Electromagnetic compatibility; Frequency; Hospitals; Medical tests; Telephone sets; Transceivers;
Conference_Titel :
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location :
Montreal, Que.
Print_ISBN :
0-7803-2475-7
DOI :
10.1109/IEMBS.1995.575311