• DocumentCode
    2990497
  • Title

    Ultrasonic and microwave irradiator for treatment of inner cavities

  • Author

    Belavskaya, S.V. ; Lisitsyna, L.I. ; Adoniev, V.G. ; Vityugov, V.A. ; Polyakov, Yu.S.

  • Author_Institution
    Novosibirsk State Tech. Univ., Russia
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    54
  • Lastpage
    56
  • Abstract
    The ultrasonic and microwave irradiator for treatment of inner cavities, fabricated on the lateral surface of urological metallic tube with embedded piezoceramic plate, is considered. An analytical model for estimation of insonation intensity is developed, and parameters for matching with lossy media are obtained. The developed mathematical model is validated by experiment
  • Keywords
    biomedical equipment; biomedical ultrasonics; hyperthermia; microwave heating; US irradiator; analytical model; embedded piezoceramic plate; inner cavities treatment; insonation intensity; lateral surface; lossy media matching; microwave irradiator; multifunctional device; urological metallic tube; Analytical models; Coaxial cables; Damping; Hydrogen; Mathematical model; Matrix converters; Medical treatment; Modems; Piezoelectric materials; Surface treatment;