DocumentCode
2990497
Title
Ultrasonic and microwave irradiator for treatment of inner cavities
Author
Belavskaya, S.V. ; Lisitsyna, L.I. ; Adoniev, V.G. ; Vityugov, V.A. ; Polyakov, Yu.S.
Author_Institution
Novosibirsk State Tech. Univ., Russia
fYear
2000
fDate
2000
Firstpage
54
Lastpage
56
Abstract
The ultrasonic and microwave irradiator for treatment of inner cavities, fabricated on the lateral surface of urological metallic tube with embedded piezoceramic plate, is considered. An analytical model for estimation of insonation intensity is developed, and parameters for matching with lossy media are obtained. The developed mathematical model is validated by experiment
Keywords
biomedical equipment; biomedical ultrasonics; hyperthermia; microwave heating; US irradiator; analytical model; embedded piezoceramic plate; inner cavities treatment; insonation intensity; lateral surface; lossy media matching; microwave irradiator; multifunctional device; urological metallic tube; Analytical models; Coaxial cables; Damping; Hydrogen; Mathematical model; Matrix converters; Medical treatment; Modems; Piezoelectric materials; Surface treatment;
fLanguage
English
Publisher
ieee
Conference_Titel
Electronic Instrument Engineering Proceedings, 2000. APEIE-2000. Volume 1. 2000 5th International Conference on Actual Problems of
Conference_Location
Novosibirsk
Print_ISBN
0-7803-5903-8
Type
conf
DOI
10.1109/APEIE.2000.913088
Filename
913088
Link To Document