Title
Voice-coil actuator control design for use in an elastance-based mock circulatory system
Author
Loh, Matthew ; Yu, Yih-Choung
Author_Institution
Dept. of Electr. & Comput. Eng., Lafayette Coll., Easton, PA, USA
fYear
2004
fDate
17-18 April 2004
Firstpage
87
Lastpage
88
Abstract
The use of a voice-coil actuator (VCA) in an elastance-based mock circulatory system (MCS) was investigated. Two VCA models, an ideal linear model and a non-linear model, were compared for use in the design of the control scheme. Data for parameter estimation was acquired by exciting the VCA with a sine wave input. VCA parameters were then identified using least-squares estimation (LSE) and curve-fitting software. LSE using the ideal linear model yielded a high RMS error and was deemed unreliable for use in controller design. Thus, the use of the non-linear model was proposed and verified in simulation. This scheme will then be implemented in hardware for further investigation.
Keywords
actuators; biocontrol; cardiovascular system; controllers; haemodynamics; least squares approximations; parameter estimation; physiological models; three-term control; curve-fitting software; elastance-based mock circulatory system; ideal linear model; least-squares estimation; nonlinear model; parameter estimation; voice-coil actuator control design; Actuators; Animals; Cardiac disease; Circulatory system; Control design; Heart; Mathematical model; Pressure control; Table lookup; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 2004. Proceedings of the IEEE 30th Annual Northeast
Print_ISBN
0-7803-8285-4
Type
conf
DOI
10.1109/NEBC.2004.1300006
Filename
1300006
Link To Document