DocumentCode
3011699
Title
nxn Stimulator and Reader Design for Retinal Prosthesis using Multiplexing Technique and CMOS Technology
Author
Talukder, Mohammad ; Siy, Pepe ; Auner, Gregory
Author_Institution
Wayne State Univ., Detroit, MI
fYear
2005
fDate
16-19 March 2005
Firstpage
410
Lastpage
413
Abstract
Serially received multiplexed pixels of an image are decoded and scanned on the probes array to stimulate the retina to reproduce the image. This paper presents a retinal stimulator (RS), which needs only six interface wires irrespective of array size. RS decodes the time-multiplexed pixels of the image and scans them on the probes array. RS doesn´t need address to route the pixels. It is a bidirectional device that can also retrieve responses from retina as a reader and sends out multiplexed readings through reverse telemetry
Keywords
biomedical optical imaging; biomedical telemetry; decoding; eye; medical image processing; prosthetics; time division multiplexing; CMOS technology; bidirectional device; image decoding; multiplexing technique; nxn stimulator; probe array; reader design; retinal prosthesis; retinal stimulator; reverse telemetry; time-multiplexed pixels; CMOS technology; Decoding; Frequency; Implants; Neural prosthesis; Neurons; Photoreceptors; Prosthetics; Retina; Wires;
fLanguage
English
Publisher
ieee
Conference_Titel
Neural Engineering, 2005. Conference Proceedings. 2nd International IEEE EMBS Conference on
Conference_Location
Arlington, VA
Print_ISBN
0-7803-8710-4
Type
conf
DOI
10.1109/CNE.2005.1419646
Filename
1419646
Link To Document