Title :
Simulation of parallel current injection for use in a vision prosthesis
Author :
Lovell, N.H. ; Dokos, S. ; Cheng, E. ; Suaning, G.J.
Author_Institution :
Graduate Sch. of Biomed. Eng., New South Wales Univ., Sydney, NSW
Abstract :
A 2D computational model was developed to simulate the distribution of voltage arising from multiple parallel current sources connected to a bipolar electrode array and immersed in a bath of physiological saline. The model was solved using a custom spectral collocation method. Pairs of electrodes placed in various locations within the bath were simulated in order to quantify return current interaction between parallel sources. A more complex electrode array geometry based on an anticipated hexagonal electrode array - to be used in a vision prosthesis - was also investigated. The guarding affect of the hexagonal array, as opposed to more traditional bipolar electrode arrangements was clearly demonstrated
Keywords :
arrays; bioelectric potentials; biomedical electrodes; physiological models; prosthetics; vision; 2D computational model; bipolar electrode array; custom spectral collocation method; hexagonal electrode array; multiple parallel current sources; parallel current injection; physiological saline; vision prosthesis; Australia; Biomedical engineering; Computational modeling; Concurrent computing; Distributed computing; Electrodes; Implants; Neural prosthesis; Prosthetics; Voltage;
Conference_Titel :
Neural Engineering, 2005. Conference Proceedings. 2nd International IEEE EMBS Conference on
Conference_Location :
Arlington, VA
Print_ISBN :
0-7803-8710-4
DOI :
10.1109/CNE.2005.1419658