DocumentCode :
3022708
Title :
A harmonic cancellation technique for an ultrasound transducer excited by a switched-mode power converter
Author :
Tang, Sai Chun ; Clement, Gregory T.
Author_Institution :
Dept. of Radiol., Harvard Med. Sch., Boston, MA
fYear :
2008
fDate :
2-5 Nov. 2008
Firstpage :
2076
Lastpage :
2079
Abstract :
The aim of this study is to evaluate the feasibility of using harmonic cancellation for a therapeutic ultrasound transducer excited by a switched-mode power converter without an additional output filter. A switching waveform without the 3rd harmonic was created by cascading two switched-mode power inverter modules at which their output waveforms were pi/3 phase-shifted from each other. The waveform and harmonic contents of the acoustic pressure generated by a 1-MHz self-focused piezoelectric transducer with and without harmonic cancellation have been evaluated. Measured results indicated that the acoustic 3rd harmonic-to-fundamental ratio at the focus was small (-48 dB) with harmonic cancellation, compared to that without harmonic cancellation (-20 dB). The measured acoustic levels of the 5th harmonic for both cases with and without harmonic cancellation were also small (-46 dB) compared to the fundamental. This study shows that it is viable to drive a piezoelectric ultrasound transducer using a switched-mode power converter without the requirement of an additional output filter in many high intensity focused ultrasound (HIFU) applications.
Keywords :
biomedical transducers; harmonics suppression; invertors; piezoelectric transducers; switched mode power supplies; switching convertors; ultrasonic focusing; ultrasonic therapy; ultrasonic transducers; HIFU; acoustic pressure generation; frequency 1 MHz; harmonic cancellation technique; high-intensity focused ultrasound application; self-focused piezoelectric transducer; switched-mode power converter; switched-mode power inverter modules; switching waveform; therapeutic ultrasound transducer; Acoustic measurements; Acoustic waves; Inverters; Piezoelectric transducers; Power harmonic filters; Power system harmonics; Switching converters; Ultrasonic imaging; Ultrasonic transducers; Ultrasonic variables measurement; HIFU; harmonic cancellation; switched-mode power converter; ultrasound transducer excitation;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium, 2008. IUS 2008. IEEE
Conference_Location :
Beijing
Print_ISBN :
978-1-4244-2428-3
Electronic_ISBN :
978-1-4244-2480-1
Type :
conf
DOI :
10.1109/ULTSYM.2008.0513
Filename :
4803459
Link To Document :
بازگشت