DocumentCode
3023123
Title
Evaluation of inline transmitter/receiver system for intravascular ultrasound imaging using Pb(Zn1/3 Nb2/3 )O3 -PbTiO3 single crystal and polyvinylidene fluoride
Author
Tanabe, Masayuki ; Okubo, Kan ; Tagawa, Norio ; Moriya, Tadashi
Author_Institution
Div. of Inf. & Commun. Syst. Eng., Tokyo Metropolitan Univ., Hino
fYear
2008
fDate
2-5 Nov. 2008
Firstpage
1765
Lastpage
1768
Abstract
Tissue harmonic imaging (THI) can get fine images, and by adopting pulse compression technique (PCT) to it, imaging with high signal-to-noise ratio can be realized. Preliminarily, we had proposed an inline transmitter/receiver system using THI and PCT for intravascular ultrasound imaging (IVUS), and examined the feasibility of the proposed system. In this study, we examined the suitable thickness of the PVDF film as not only a matching layer but also a receiver for harmonic components.
Keywords
biomedical transducers; biomedical ultrasonics; blood vessels; lead compounds; polymer films; thin film devices; ultrasonic transducers; IVUS; PCT; PVDF film; Pb(Zn0.33Nb0.66)O3-PbTiO3; THI; inline transmitter-receiver system; intravascular ultrasound imaging; matching layer; polyvinylidene fluoride; pulse compression technique; single crystal; tissue harmonic imaging; Bandwidth; Computational modeling; High-resolution imaging; Optical films; Piezoelectric films; Pressure measurement; Pulse compression methods; Thickness measurement; Transmitters; Ultrasonic imaging; IVUS; PVDF; PZN-PT; chirp; component; pulse compression; tissue harmonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2008. IUS 2008. IEEE
Conference_Location
Beijing
Print_ISBN
978-1-4244-2428-3
Electronic_ISBN
978-1-4244-2480-1
Type
conf
DOI
10.1109/ULTSYM.2008.0433
Filename
4803476
Link To Document