• DocumentCode
    3024623
  • Title

    Design and development of a 3-axis MRI-compatible force sensor

  • Author

    Tan, U-Xuan ; Yang, Bo ; Gullapalli, Rao ; Desai, Jaydev P.

  • Author_Institution
    Dept. of Mech. Eng., Univ. of Maryland, College Park, MD, USA
  • fYear
    2010
  • fDate
    3-7 May 2010
  • Firstpage
    2586
  • Lastpage
    2591
  • Abstract
    Magnetic resonance imaging (MRI) has been gaining popularity over standard imaging modalities like ultrasound and CT because of its ability to provide excellent soft-tissue contrast. However, due to the working principle of MRI, a number of conventional force sensors are not compatible. One popular solution is to develop a fiber-optic force sensor. However, the measurements along the principal axes of a number of these force sensors are highly cross-coupled. One of the objectives of this paper is to minimize this coupling effect. In addition, this paper describes the design of an elastic frame structure that is obtained systematically by an algorithm and not purely based on design intuition. We used a topology optimization technique, which has two major advantages: 1) aids engineers in design when given a constrained boundary, and 2) optimize the displacement amplification, which will in turn increase stiffness, bandwidth, and improve sensing resolution. To ensure that the frames are linked from the input to output, a solution for topology optimization is proposed. The sensor is then fabricated using plastic material (ABS) as it is one of the ideal material for MRI environment. However, the hysteresis effect seen in the displacement-load graph of plastic materials is known to affect the accuracy. Hence, this paper also proposes modeling and addressing this hysteretic effect using Prandtl-Ishlinskii play operators. Finally, experiments are conducted to evaluate the sensor´s performance, as well as its compatibility in MRI under continuous imaging.
  • Keywords
    biomedical MRI; fibre optic sensors; force sensors; medical control systems; optimisation; topology; MRI-compatible force sensor; cross-coupling effect; displacement amplification; elastic frame structure design; fiber-optic force sensor; magnetic resonance imaging; soft-tissue contrast; topology optimization technique; Algorithm design and analysis; Computed tomography; Conducting materials; Constraint optimization; Design optimization; Force sensors; Magnetic resonance imaging; Plastics; Topology; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Robotics and Automation (ICRA), 2010 IEEE International Conference on
  • Conference_Location
    Anchorage, AK
  • ISSN
    1050-4729
  • Print_ISBN
    978-1-4244-5038-1
  • Electronic_ISBN
    1050-4729
  • Type

    conf

  • DOI
    10.1109/ROBOT.2010.5509761
  • Filename
    5509761