Title :
Optimization of beams with nonspherical extended depths of focus for reconfigurable 2D arrays
Author :
Hooi, Fong Ming ; Carson, Paul L. ; Thomenius, Kai ; Fisher, Rayette A.
Author_Institution :
Dept. of Radiol., Univ. of Michigan, Ann Arbor, MI
Abstract :
The flexibility of reconfigurable 2D arrays allows us to utilize various beamforming techniques with a limited number of channels. We investigated a different beamforming technique incorporating spherical and axicon focusing to attain an extended range of focus in the transmit beam. In this paper, we develop and optimize this hybrid beamforming technique for balance between range and image quality. A local minimization technique is discussed and four apertures chosen to cover a 4.5 cm depth for breast imaging were compared to pure spherical beams via PSF and CNR calculations. Improvements were statistically significant at focal depth extremes.
Keywords :
acoustic arrays; acoustic signal processing; array signal processing; biomedical ultrasonics; mammography; minimisation; ultrasonic focusing; CNR calculations; PSF calculations; axicon focusing; beam optimization; beamforming techniques; breast imaging; nonspherical extended focus depths; reconfigurable 2D arrays; spherical focusing; Acoustic arrays; Acoustic beams; Acoustic imaging; Apertures; Array signal processing; Biomedical acoustics; Biomedical imaging; Breast; Focusing; Optical imaging; Array signal processing; acoustic arrays; biomedical acoustics; circular arrays;
Conference_Titel :
Ultrasonics Symposium, 2008. IUS 2008. IEEE
Conference_Location :
Beijing
Print_ISBN :
978-1-4244-2428-3
Electronic_ISBN :
978-1-4244-2480-1
DOI :
10.1109/ULTSYM.2008.0305