    Ultrasound computer tomography using synthetic aperture focusing technique is sensitive to phase aberration errors. We propose a signal preprocessing method for coherent imaging, which can take the system specific phase aberration into account. The method detects local maxima in the envelope of a matched filtered signal and convolutes the result with a truncated difference of sinc-functions, where the main lobe is scaled to represent the phase aberration. For our system we could show that reflectivity images reconstructed such pre-processed signals are approximately within a factor two of the ideal resolution, while increasing the contrast by a factor of 30.
  • Keywords
    aberrations; acoustic tomography; biomedical ultrasonics; computerised tomography; image reconstruction; image resolution; matched filters; medical image processing; ultrasonic focusing; 3D ultrasound computer tomography image; breast cancer imaging; coherent imaging; ideal resolution; image reconstruction; matched filtered signal; phase aberration error; signal pre-processing; synthetic aperture focusing technique; Computer errors; Envelope detectors; Focusing; Image reconstruction; Image resolution; Matched filters; Phase detection; Reflectivity; Tomography; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2008. IUS 2008. IEEE
  • Conference_Location
    Beijing
  • Print_ISBN
    978-1-4244-2428-3
  • Electronic_ISBN
    978-1-4244-2480-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2008.0205
  • Filename
    4803642