    Cuffless and noninvasive estimation of blood pressure based on a wavelet transform approach

  • Author

    Lee, C.M. ; Zhang, Y.T.

  • Author_Institution
    Joint Res. Center for Biomed. Eng., Chinese Univ. of Hong Kong, China
  • fYear
    2003
  • fDate
    20-22 Oct. 2003
  • Firstpage
    148
  • Lastpage
    149
  • Abstract
    In this paper, pulse transit time (PTT) was calculated based on wavelet transform modulus maxima (WTMM), for the estimation of blood pressure (BP). The mean and standard deviation of the differences between the estimated BP values and the values obtained from the standard BP meter were -2.6256 mmHg and 6.7459 mmHg fur systolic blood pressure, and -0.7901 mmHg and 6.1777 mmHg for diastolic blood pressure respectively. The results show that the WTMM-based BP estimation approach is quite promising.
  • Keywords
    blood pressure measurement; electrocardiography; medical signal processing; plethysmography; statistical analysis; wavelet transforms; -0.7901 mmHg; -2.6256 mmHg; 6.1777 mmHg; ECG; blood pressure; blood pressure estimation; cuffless estimation; diastolic blood pressure; mean; noninvasive estimation; photoplethysmogram; pulse transit time; standard deviation; systolic blood pressure; wavelet transform approach; wavelet transform modulus maxima; Arteries; Biomedical engineering; Blood pressure; Discrete wavelet transforms; Ear; Electrocardiography; Fingers; Foot; Frequency; Wavelet transforms;