Title :
Combining Surface Treatments With Shallow Slots to Improve the Spatial Resolution Performance of Continuous, Thick LYSO Detectors for PET
Author :
Kaul, Manohar ; Surti, Suleman ; Karp, Joel S.
Author_Institution :
Dept. of Phys., Univ. of Pennsylvania, Philadelphia, PA, USA
Abstract :
Positron emission tomography (PET) detectors based on continuous scintillation crystals can achieve very good performance and have a number of practical advantages compared to detectors based on a pixelated array of crystals. Our goal is to develop a thick continuous detector with high energy and spatial resolution, along with high γ -photon capture efficiency. We examine the performance of two crystal blocks: a 46 × 46 × 14 mm3 and a 48×48×25 mm3 block of LYSO (Lutetium Yttrium Orthosilicate). Using Maximum Likelihood (ML) positioning based upon the light response function (LRF) in the 14 mm thick crystal, we measure a spatial resolution of 3 mm in the central region of the crystal with degradation near the edges due to reflections off the crystal sides. We also show that we can match the spatial resolution achieved using a 14 mm thick crystal by using a 25 mm thick crystal with slots cut into the gamma entrance surface to narrow the LRF. We also find that we can improve the spatial resolution performance near the detector edges by reducing the reflectivity of the crystal sides, albeit with some loss in energy resolution.
Keywords :
chemical sensors; lutetium compounds; maximum likelihood estimation; positron emission tomography; surface treatment; yttrium compounds; γ-photon capture efficiency; LRF; Lu2-xYxSiO5; ML positioning; PET detectors; continuous detector; continuous scintillation crystals; gamma entrance surface; light response function; lutetium yttrium orthosilicate; maximum likelihood positioning; pixelated array; size 14 mm; size 25 mm; spatial resolution performance; surface treatments; Anodes; Crystals; Detectors; Positron emission tomography; Spatial resolution; Surface treatment; Thickness measurement; Continuous crystals; depth of interaction (DOI); maximum likelihood positioning; positron emission tomography (PET); slotted LYSO; spatial resolution;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2013.2240315