• DocumentCode
    3044819
  • Title

    Wireless EEG system with real time impedance monitoring and active electrodes

  • Author

    Patki, Shrishail ; Grundlehner, Bernard ; Verwegen, A. ; Mitra, Subhasish ; Xu, Jie ; Matsumoto, Akiyoshi ; Yazicioglu, Refet Firat ; Penders, Julien

  • Author_Institution
    Holst Centre, imec-nl, Netherlands
  • fYear
    2012
  • fDate
    28-30 Nov. 2012
  • Firstpage
    108
  • Lastpage
    111
  • Abstract
    In this paper, we present a miniaturized (<;6cm3) and low noise (60nV/√Hz) wireless EEG sensor node with active electrodes and simultaneous electrode tissue impedance (ETI) monitoring. The added benefit of the active electrodes and continuous ETI monitoring is quantified in terms of susceptibility against power line interference and cable motion artefacts. The sensor node is benchmarked against a reference system for similarity measures in EEG frequency response in the relevant bandwidth. Applications of this prototype are foreseen in the clinical, lifestyle and entertainment domains.
  • Keywords
    biological tissues; biomedical electrodes; electroencephalography; wireless sensor networks; EEG frequency response; active electrodes; cable motion artefacts; continuous ETI monitoring; low noise wireless EEG sensor node; power line interference; real time impedance monitoring; reference system; relevant bandwidth; sensor node; simultaneous electrode tissue impedance monitoring; wireless EEG system; Electrodes; Electroencephalography; Impedance; Monitoring; Noise; Wireless communication; Wireless sensor networks;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2012 IEEE
  • Conference_Location
    Hsinchu
  • Print_ISBN
    978-1-4673-2291-1
  • Electronic_ISBN
    978-1-4673-2292-8
  • Type

    conf

  • DOI
    10.1109/BioCAS.2012.6418408
  • Filename