Title :
Evaluation of Agilent 3458A Time Jitter Performance
Author :
Lapuh, Rado ; Voljc, Bostjan ; Lindic, Matjaz
Author_Institution :
Slovenian Inst. of Quality & Metrol., Ljubljana, Slovenia
Abstract :
This paper describes the mechanism of an Agilent 3458A sampling time jitter present when external triggering is used at synchronous two-channel sampling. Based on measurements, it is shown that the master digital multimeter (DMM) adds approximately 3.2 ns of its own time jitter while triggering the slave DMM and that the effective time jitter for externally triggered DMM remains at around 7 ns, well below ±50-ns maximum jitter as specified by the manufacturer.
Keywords :
digital multimeters; sampling methods; timing jitter; trigger circuits; Agilent 3458A sampling time jitter; digital multimeter; slave DMM; synchronous two-channel sampling; triggering; Clocks; Frequency measurement; Jitter; Metrology; Signal to noise ratio; Timing; Estimation; measurement; noise; quantization; sampling; signal-to-noise ratio (SNR); synchronization; time jitter; uncertainty; uncertainty.;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2015.2408793