    K-space and image space combination for motion artifact correction in multicoil multishot diffusion weighted imaging

  • Author

    Van, Anh T. ; Karampinos, Dimitrios C. ; Georgiadis, John G. ; Sutton, Bradley P.

  • Author_Institution
    Electrical and Computer Engineering Department, University of Illinois at Urbana-Champaign, USA
  • fYear
    2008
  • fDate
    20-25 Aug. 2008
  • Firstpage
    1675
  • Lastpage
    1678
  • Abstract
    Motion during diffusion encodings leads to phase errors in different shots of a multishot acquisition. Phase differences in k-space data among shots result in phase cancellation artifacts in the reconstructed image. Due to aliasing of the phase from under-sampled regions of the shot, correction of the phase error using direct low-resolution phase subtraction is incomplete. We introduce a new k-space and image-space combination (KICT) method for motion artifacts cancellation that avoids incomplete phase error correction. Further, the method preserves the phase of the object, which is important for parallel imaging applications.
  • Keywords
    Degradation; Encoding; Error correction; Image coding; Image reconstruction; Image resolution; Microstructure; Motion analysis; Phase estimation; Reconstruction algorithms; Biomedical Engineering; Brain; Diffusion Magnetic Resonance Imaging; Humans; Image Processing, Computer-Assisted; Models, Statistical; Motion; Sensitivity and Specificity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-1814-5
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2008.4649497
  • Filename
    4649497