DocumentCode
3069516
Title
A ultra high field multi-element transceive volume array for small animal MRI
Author
Weber, Ewald ; Li, Bing Keong ; Liu, Feng ; Crozier, Stuart
Author_Institution
University of Queensland, School of Information Technology and Electrical Engineering, Australia
fYear
2008
fDate
20-25 Aug. 2008
Firstpage
2039
Lastpage
2042
Abstract
In this work, a dedicated, 9.4T shielded 8-element transceive volume-array for large rat MRI applications is designed and the prototype constructed. Angularly oriented radiating structured coil elements are used. It is shown that by orienting the radiating conductors of the coil element angularly, the RF penetration depth can be increased and the mutual coupling effect can be regulated to minimum. In addition, experimental results show that the prototype can produce homogenous B1 field and is well suited for the Transmit SENSE application and the GRAPPA parallel imaging technique.
Keywords
Animal structures; Coils; Conductors; Dielectrics; Information technology; Magnetic resonance imaging; Mutual coupling; Nonuniform electric fields; Prototypes; Radio frequency; Algorithms; Animals; Equipment Design; Magnetic Resonance Imaging; Rats; Signal Processing, Computer-Assisted; Software; Transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location
Vancouver, BC
ISSN
1557-170X
Print_ISBN
978-1-4244-1814-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2008.4649592
Filename
4649592
Link To Document