• DocumentCode
    3071252
  • Title

    H-reflex recorded by multi-pad EMG electrodes

  • Author

    Kojic, V. ; Miljkovic, N. ; Malesevic, Nebojsa ; Popovic, Dejan B.

  • Author_Institution
    Sch. of Electr. Eng., Univ. of Belgrade, Belgrade, Serbia
  • fYear
    2012
  • fDate
    20-22 Sept. 2012
  • Firstpage
    119
  • Lastpage
    122
  • Abstract
    We recorded H-reflex using 4 × 4 multi-pad EMG electrodes on muscle Soleus from 4 healthy subjects. The results indicate that the new proposed method of recording H-reflex provides more information than standard bipolar or unipolar EMG recording, since it enables spatial and temporal parameters assessment. We have developed an algorithm which can automatically select the H-reflex with maximum peak to peak amplitude. We also examined the H-reflex propagation by the use of EMG topography map based on the latency.
  • Keywords
    biomedical electrodes; electromyography; EMG topography map; H-reflex; bipolar EMG recording; multipad EMG electrodes; muscle Soleus; spatial parameters assessment; temporal parameters assessment; unipolar EMG recording; Educational institutions; Electrodes; Electromyography; Muscles; Surface topography; USA Councils; EMG; H-reflex; latency; multi-pad electrode; topography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Neural Network Applications in Electrical Engineering (NEUREL), 2012 11th Symposium on
  • Conference_Location
    Belgrade
  • Print_ISBN
    978-1-4673-1569-2
  • Type

    conf

  • DOI
    10.1109/NEUREL.2012.6419981
  • Filename
    6419981