We studied the temperature gradient of radio-frequency energy delivered intra-murally in-vivo. The objective was to compare the measured temperatures obtained in living myocardium with analytical estimations of temperature gradients. Closed loop temperature control was implemented to achieve set temperatures of 70, 80, 90 and 100°C at the delivering electrode. The analytic solution was found to closely follow the measured data. We conclude that the developed solution may be used to accurately predict temperature gradients and hence lesion size
Keywords
bioelectric phenomena; biothermics; cardiology; closed loop systems; muscle; radiofrequency heating; surgery; temperature control; temperature distribution; 100 degC; 70 degC; 80 degC; 90 degC; actual intramyocardial temperature gradients; analytical estimations; closed loop temperature control; delivering electrode; in-vivo; intramural radio-frequency ablation; lesion size; living myocardium; malignant ventricular tachycardia; radio-frequency energy; surgery; Cardiology; Electrodes; Impedance; Lesions; Myocardium; Needles; Radio frequency; Temperature control; Temperature measurement; Temperature sensors;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1995., IEEE 17th Annual Conference
Conference_Location
Montreal, Que.
Print_ISBN
0-7803-2475-7
Type
conf
DOI
10.1109/IEMBS.1995.575105
Filename
575105
Link To Document