    Temperature distribution under cooled electrodes during radiofrequency catheter ablation

  • Author

    Panescu, Dorin ; Fleischman, Sidney D. ; Whayne, James G. ; Swanson, David K.

  • Author_Institution
    EP Technol. Inc., Sunnyvale, CA, USA
  • Volume
    1
  • fYear
    1995
  • fDate
    20-25 Sep 1995
  • Firstpage
    299
  • Abstract
    Larger lesions were created when the electrode of an RF ablation system was actively cooled below blood temperature. To keep the maximal tissue temperature at about 90°C, significantly more RF power had to be delivered. Because the hottest tissue region moved farther from the electrode surface, it was more difficult to predict the adequate tissue heating. Therefore, the probability of overheating the tissue and producing micro-explosions increased
  • Keywords
    bioelectric phenomena; cooling; finite element analysis; hyperthermia; radiofrequency heating; temperature distribution; 3-D finite element model; 90 C; adequate tissue heating; blood temperature; cooled electrodes; electric heating; hottest tissue region; larger lesions; maximal tissue temperature; micro-explosions; radiofrequency catheter ablation; therapeutic technique; tissue overheating; Blood; Boundary conditions; Cardiac tissue; Catheters; Electrodes; Iron; Lesions; Radio frequency; Temperature distribution; Thermistors;