Title :
Real-time cardiac MRI without triggering, gating, or breath holding
Author :
Brinegar, Cornelius ; Wu, Yi-Jen L. ; Foley, Lesley M. ; Hitchens, T. Kevin ; Ye, Qing ; Ho, Chien ; Liang, Zhi-Pei
Author_Institution :
Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, 1406 West Green Street, 61801, USA
Abstract :
State-of-the-art cardiac MRI can perform real-time 2D scans without cardiac triggering during a single breath hold; however, real-time cardiac MRI in rats is difficult due to the high heart rate (330 bpm) and presence of respiratory motion. These challenges are overcome by using a dynamic imaging method based on Partially Separable Function (PSF) theory with an acceleration factor of 256. This paper demonstrates that this method can be used in the study of transplanted rat hearts for both anatomical and perfusion applications. The study was carried out with a 200 μm in-plane resolution with a 17.2 msec temporal resolution, and the results show improved spatial resolution (2x) and reduced acquisition time (3x) relative to Electrocardiogram (ECG) triggered, respiratory gated cine imaging.
Keywords :
Acceleration; Electrocardiography; Equations; Heart rate; High-resolution imaging; Image resolution; Immune system; Magnetic resonance imaging; Rats; Spatial resolution; Animals; Contrast Media; Electrocardiography; Gadolinium; Heart; Heart Rate; Heart Transplantation; Image Enhancement; Image Processing, Computer-Assisted; Magnetic Resonance Imaging; Models, Statistical; Perfusion; Rats; Reference Values; Reproducibility of Results;
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2008.4649931