  • Abstract
    To develop phased array surface coils, noise correlation between coil elements should be carefully considered. Here, an analytical expression has been derived which indicates the relation between noise correlation and transmission coefficient of any arbitrary two-port network. According to this result, it is clear that the transmission coefficient describes completely the noise correlation of the coils. To achieve minimum noise correlation for a coil system, one only needs to reduce the transmission coefficient s21 between coil 1 and coil 2
  • Keywords
    biomedical NMR; biomedical equipment; coils; noise; MRI; analytical expression; arbitrary two-port network; coil elements; magnetic resonance imaging; medical diagnostic imaging; noise correlation; phased array surface coils; surface coils; transmission coefficient; Circuit noise; Coils; Electrical resistance measurement; Equivalent circuits; Impedance; Magnetic noise; Noise measurement; Noise reduction; Phased arrays; Scattering parameters;