Title :
Preliminary imaging results of magnetic induction tomography based on physical phantom
Author :
Liu, Ruigang ; Li, Ye ; You, Fusheng ; Shi, Xuetao ; Fu, Feng ; Dong, Xiuzhen
Author_Institution :
Faculty of the Biomedical Engineering, the Fourth Military Medical University, Xi´´an, China
Abstract :
Magnetic induction tomography (MIT) is a kind of contactless and noninvasive method to reconstruct the conductivity distribution in a cross-section. We developed a data acquisition system with one exciting coil and 15 measuring coils. By rotating the object placing on a circle plate, the 16*15 measurement data were obtained. The Newton-Raphson algorithm was applied to reconstruct the conductivity distribution. To the conductivity perturbation objects in air field, we made the difference images of MIT. The images were able to demonstrate the location of the objects and identify the conductivity difference between 0.84 S/m and 1.26 S/m, which are in the range of biological tissue.
Keywords :
Biomedical measurements; Coils; Conductivity measurement; Current measurement; Data acquisition; Imaging phantoms; Impedance; Magnetic field measurement; Rotation measurement; Tomography; Algorithms; Electric Conductivity; Electric Impedance; Equipment Design; Finite Element Analysis; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Magnetics; Models, Statistical; Software; Tomography;
Conference_Titel :
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location :
Vancouver, BC
Print_ISBN :
978-1-4244-1814-5
Electronic_ISBN :
1557-170X
DOI :
10.1109/IEMBS.2008.4650227