DocumentCode
3081762
Title
Bandwidth measurement of the conductance catheter system
Author
Chen, Chieh-En ; Wei, Chia-Ling
Author_Institution
Department of Electrical Engineering, National Cheng Kung University, Tainan, Taiwan
fYear
2008
fDate
20-25 Aug. 2008
Firstpage
4859
Lastpage
4862
Abstract
Conductance catheter system is used to estimate real-time ventricular volume by measuring the time-varying ventricular conductance/admittance. However, the system is generally calibrated only with known resistors, while neither the frequency response nor the bandwidth of the system is calibrated and measured. The main difficulty of measuring its bandwidth is that the sensed signal of the conductance catheter system, which can be viewed as the input of the system, is a modulated signal, rather than a typical sin wave. Therefore, its bandwidth cannot be measured by typical frequency response analyzers, since they are designed for pure sin-wave input. The waveform of the sensed signal is analyzed. Moreover, a waveform simulator designed to mimic the sensed signal is presented. It can be used to measure the bandwidth of conductance catheter system.
Keywords
Admittance measurement; Bandwidth; Catheters; Frequency measurement; Frequency response; Real time systems; Resistors; Signal analysis; Time varying systems; Volume measurement; Cardiac Volume; Cardiography, Impedance; Equipment Design; Equipment Failure Analysis; Heart Catheterization; Humans; Reproducibility of Results; Sensitivity and Specificity;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location
Vancouver, BC
ISSN
1557-170X
Print_ISBN
978-1-4244-1814-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2008.4650302
Filename
4650302
Link To Document