• DocumentCode
    3081762
  • Title

    Bandwidth measurement of the conductance catheter system

  • Author

    Chen, Chieh-En ; Wei, Chia-Ling

  • Author_Institution
    Department of Electrical Engineering, National Cheng Kung University, Tainan, Taiwan
  • fYear
    2008
  • fDate
    20-25 Aug. 2008
  • Firstpage
    4859
  • Lastpage
    4862
  • Abstract
    Conductance catheter system is used to estimate real-time ventricular volume by measuring the time-varying ventricular conductance/admittance. However, the system is generally calibrated only with known resistors, while neither the frequency response nor the bandwidth of the system is calibrated and measured. The main difficulty of measuring its bandwidth is that the sensed signal of the conductance catheter system, which can be viewed as the input of the system, is a modulated signal, rather than a typical sin wave. Therefore, its bandwidth cannot be measured by typical frequency response analyzers, since they are designed for pure sin-wave input. The waveform of the sensed signal is analyzed. Moreover, a waveform simulator designed to mimic the sensed signal is presented. It can be used to measure the bandwidth of conductance catheter system.
  • Keywords
    Admittance measurement; Bandwidth; Catheters; Frequency measurement; Frequency response; Real time systems; Resistors; Signal analysis; Time varying systems; Volume measurement; Cardiac Volume; Cardiography, Impedance; Equipment Design; Equipment Failure Analysis; Heart Catheterization; Humans; Reproducibility of Results; Sensitivity and Specificity;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
  • Conference_Location
    Vancouver, BC
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-1814-5
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2008.4650302
  • Filename
    4650302