• DocumentCode
    3113527
  • Title

    Capacitive Electrodes in Electroencepahlography

  • Author

    Von Ellenrieder, Nicolas ; Spinelli, Enrique ; Muravchik, Carlos H.

  • Author_Institution
    Area Dept. Electrotecnia, Univ. Nacional de La Plata
  • fYear
    2006
  • fDate
    Aug. 30 2006-Sept. 3 2006
  • Firstpage
    1126
  • Lastpage
    1129
  • Abstract
    We present a forward problem formulation for computing biopotentials measured with dry or capacitive electrodes. This formulation is not quasistatic and has mixed boundary conditions. Our results show that simple approximations to the measurements based on capacitive coupling are adequate in most situations. We study the range of validity and errors committed in the EEG forward and inverse problems when using this approximation
  • Keywords
    bioelectric potentials; biomedical electrodes; biomedical measurement; capacitive sensors; electroencephalography; inverse problems; biopotential computation; capacitive coupling; capacitive electrodes; dry electrodes; electroencephalography; forward problems; inverse problems; mixed boundary conditions; Biological materials; Boundary conditions; Conducting materials; Dielectrics and electrical insulation; Electric potential; Electrodes; Electroencephalography; Inverse problems; Magnetic heads; Skin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
  • Conference_Location
    New York, NY
  • ISSN
    1557-170X
  • Print_ISBN
    1-4244-0032-5
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2006.260226
  • Filename
    4461954