Measurement of microphone parameters based on high-pressure calibration system
Author :
Feng Niu ; Aijian Zheng ; Haijiang Zhu ; Ping Yang ; Longbiao He
Author_Institution :
Div. of Mech. & Acoust., Nat. Inst. of Metrol., Beijing, China
fYear :
2012
fDate :
5-8 Aug. 2012
Firstpage :
687
Lastpage :
691
Abstract :
This paper presents a method for measuring a few microphone parameters based on the high-pressure calibration system. The high pressure calibration apparatus and method are first given. Then the linearity-deviations of three microphones are measured, the uncertainties of three microphones are estimated, and the dynamic range limits of three microphones at high sound pressure levels (SPL) are measured. In this study, the range of the linearity-deviations is from -0.02dB to +0.02dB, and that of the uncertainties is less 0.05dB. The dynamic range limit of three microphones is 149dB, 147dB and 162dB, respectively. Compared with the manufacturer parameters, experimental results are consistent with manufacturer data.
Keywords :
acoustic intensity measurement; calibration; measurement uncertainty; microphones; SPL; dynamic range limits; high-pressure calibration system; linearity-deviations; measurement uncertainties; microphone parameter measurement; sound pressure levels; Calibration; Distortion measurement; Dynamic range; Frequency measurement; Microphones; Pressure measurement; Uncertainty; Calibration System; Dynamic Range Limit; High Pressure Level;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Mechatronics and Automation (ICMA), 2012 International Conference on
Conference_Location :
Chengdu
Print_ISBN :
978-1-4673-1275-2
Type :
conf
DOI :
10.1109/ICMA.2012.6283225
Filename :
6283225
Link To Document :
بازگشت