• DocumentCode
    3121195
  • Title

    Biosensor multiwell microplate with self-reference capability

  • Author

    Cunningham, B. ; Baird, C. ; Li, P. ; Schulz, S. ; Albertson, K. ; Gerstenmaier, J.

  • Author_Institution
    SRU Biosystems, Woburn, MA, USA
  • fYear
    2004
  • fDate
    24-27 Oct. 2004
  • Firstpage
    806
  • Abstract
    The ability to remove sources of measurement error accurately and easily is one of the keys to a high-resolution biodetection system. The ability of a guided mode resonant filter (GMRF) biosensor structure to confine the lateral propagation of photons tightly is exploited to demonstrate a sensor surface upon which active and reference regions can be located in close physical proximity. Detection systems with two-point reference correction and an imaging-based reference method are demonstrated.
  • Keywords
    biosensors; measurement errors; microsensors; optical filters; optical information processing; optical sensors; biosensor multiwell microplate; guided mode resonant filter biosensor structure; high-resolution biodetection system; imaging-based reference method; lateral photon propagation; measurement error source removal; optical biosensors; self-reference capability; two-point reference correction; Biosensors; Error correction; Image sensors; Optical sensors; Optical surface waves; Optoelectronic and photonic sensors; Pixel; Resonance; Surface waves; Temperature sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2004. Proceedings of IEEE
  • Print_ISBN
    0-7803-8692-2
  • Type

    conf

  • DOI
    10.1109/ICSENS.2004.1426291
  • Filename
    1426291