  • Abstract
    Since several years, the number of total hip arthroplasty revision surgeries is substantially growing. One of the main reasons for this procedure to become necessary is the loosening or damage of the prothesis, which is facilitated by bone necrosis at the implant-bone interface. Electrostimulation is one promising technique, which can accelerate the growth of bone cells and, therefore, enhance the anchorage of the implant to the bone. We present computational models of an electrostimulative total hip revision system to enhance bone regeneration. In this study, the influence of uncertainty in the conductivity of bone tissue on the electric field strength and the beneficial stimulation volume for an optimized electrode geometry and arrangement is investigated. The generalized polynomial chaos technique is used to quantify the uncertainty in the stimulation volumes with respect to the uncertain conductivity of cancellous bone, bone marrow, and bone substitute, which is used to fill defective areas. The results suggest that the overall beneficial stimulation areas are only slightly sensitive to the uncertainty in conductivity of bone tissue. However, in the proximity of tissue boundaries, larger uncertainties, especially in the transition between beneficial and understimulation areas, can be expected.
  • Keywords
    bioelectric phenomena; biomedical electrodes; bone; cellular biophysics; chaos; electrical conductivity; optimisation; polynomials; prosthetics; surgery; beneficial stimulation areas; bone cell growth; bone marrow; bone necrosis; bone regeneration; bone substitute; bone tissue conductivity; cancellous bone; electric field strength; electrostimulation; generalized polynomial chaos technique; implant-bone interface; optimized electrode geometry; optimized electrostimulative hip revision system; prothesis damage; prothesis loosening; tissue boundaries; total hip arthroplasty revision surgeries; uncertainty; understimulation areas; Bones; Computational modeling; Conductivity; Electrodes; Hip; Implants; Uncertainty; Electrical Stimulation; Electrical stimulation; Finite Element Method; Multi-Objective Optimization; Total Hip Arthroplasty Revision; Uncertainty Quantification; finite-element method; multiobjective optimization; total hip arthroplasty (THA) revision; uncertainty quantification;
  • fLanguage
    English
  • Journal_Title
    Biomedical and Health Informatics, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    2168-2194
  • Type

    jour

  • DOI
    10.1109/JBHI.2015.2423705
  • Filename
    7088560