DocumentCode :
3128886
Title :
An automatic calibration system for anechoic chamber
Author :
Zhong Bo ; Wang Jitong ; Niu Feng ; He Longbiao ; Li Xuebin
Author_Institution :
Div. of Mech. & Acoust., Nat. Inst. of Metrol., Beijing, China
fYear :
2012
fDate :
5-8 Aug. 2012
Firstpage :
1313
Lastpage :
1317
Abstract :
The importance of automatic calibration for anechoic chamber or room is described and the problems of existing calibration systems are analyzed at the beginning of this paper. Then a novel automatic calibration system is presented. The hardware of the system contains three key components which are the microphone position unit, sound level measurement unit and sound output unit. All these components are controlled by the controlling program and the measurement data can be analyzed automatically by the analyzing program so that the final results can be obtained. All these programs are developed by ourselves according to the standard ISO 3745. The system can realize automatic calibration for the anechoic chamber including microphone movement, sound output intensity adjustment, sound level measurement and data processing. In addition, the system can generate final reports automatically with numerical and graphic formats. The experiment results show that our system has the features of high-precision, high-efficiency, easy-operation, and good-universality and it is well competent for routine calibration work for anechoic chamber.
Keywords :
acoustic intensity measurement; anechoic chambers (acoustic); architectural acoustics; calibration; measurement standards; microphones; analyzing program; anechoic chamber; automatic calibration system; automatic report generation; controlling program; data processing; graphic formats; measurement data; microphone movement; microphone position unit; numerical formats; room; sound level measurement unit; sound output intensity adjustment; sound output unit; standard ISO 3745; Acoustic measurements; Acoustics; Anechoic chambers; Calibration; Frequency measurement; Microphones; Pressure measurement; acoustic metrology; anechoic chamber; automatic calibration; free-sound-field;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Mechatronics and Automation (ICMA), 2012 International Conference on
Conference_Location :
Chengdu
Print_ISBN :
978-1-4673-1275-2
Type :
conf
DOI :
10.1109/ICMA.2012.6284326
Filename :
6284326