DocumentCode :
3138711
Title :
Vibration suppression of mobile medical imaging system for diagnosis using zero vibration control
Author :
Kim, Duckjune ; Lee, Kwang-Hee ; Ji, Sang-Hoon ; Shon, Woong-Hee
Author_Institution :
Robot Convergence Res. Center, Korea Inst. of Ind. Technol., Ansan, South Korea
fYear :
2011
fDate :
19-21 Dec. 2011
Firstpage :
579
Lastpage :
583
Abstract :
This paper presents a zero vibration control for mobile medical robot system when the gantry is rotated. ISO center needs to maintain its position while gantry is rotated to take multiple X-ray images for reconstructed 3D CT image. The gantry includes X-ray source and detector module causing unbalanced weight of rotor and the gantry is support in cantilever manner by support structure. This fact causes vibration of gantry in vertical axis, which is harmful for high quality of 3D CT image. In this paper, the manipulator is operated to give impulses to the system to reduce vibration and it is validated by experimental results.
Keywords :
X-ray imaging; cantilevers; computerised tomography; image reconstruction; manipulators; medical image processing; medical robotics; mobile robots; patient diagnosis; rotors; vibration control; 3D CT image reconstruction; ISO center; X-ray images; X-ray source; cantilever manner; detector module; gantry; manipulator; medical diagnosis; mobile medical imaging system; mobile medical robot system; rotor; unbalanced weight; vertical axis; vibration suppression; zero vibration control; Computed tomography; Manipulators; Mobile communication; Rotors; Three dimensional displays; Vibration control; Vibrations;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Control and Automation (ICCA), 2011 9th IEEE International Conference on
Conference_Location :
Santiago
ISSN :
1948-3449
Print_ISBN :
978-1-4577-1475-7
Type :
conf
DOI :
10.1109/ICCA.2011.6138020
Filename :
6138020
Link To Document :
بازگشت