Abstract
This paper presents a zero vibration control for mobile medical robot system when the gantry is rotated. ISO center needs to maintain its position while gantry is rotated to take multiple X-ray images for reconstructed 3D CT image. The gantry includes X-ray source and detector module causing unbalanced weight of rotor and the gantry is support in cantilever manner by support structure. This fact causes vibration of gantry in vertical axis, which is harmful for high quality of 3D CT image. In this paper, the manipulator is operated to give impulses to the system to reduce vibration and it is validated by experimental results.
Keywords
X-ray imaging; cantilevers; computerised tomography; image reconstruction; manipulators; medical image processing; medical robotics; mobile robots; patient diagnosis; rotors; vibration control; 3D CT image reconstruction; ISO center; X-ray images; X-ray source; cantilever manner; detector module; gantry; manipulator; medical diagnosis; mobile medical imaging system; mobile medical robot system; rotor; unbalanced weight; vertical axis; vibration suppression; zero vibration control; Computed tomography; Manipulators; Mobile communication; Rotors; Three dimensional displays; Vibration control; Vibrations;
fLanguage
English
Publisher
ieee
Conference_Titel
Control and Automation (ICCA), 2011 9th IEEE International Conference on
Conference_Location
Santiago
ISSN
1948-3449
Print_ISBN
978-1-4577-1475-7
Type
conf
DOI
10.1109/ICCA.2011.6138020
Filename
6138020
Link To Document